Measuring pH with Electrodes

Want
voltage (potential difference, E) between electrodes to be a f(pH)
- pH sensitive (glass electrode)
has E = f(pH of soln)
- reference electrode (calomel)
has Ereference = pH independent = constant
- E = EpH electrode
- Ereference = f(pH of soln)
Why a glass
electrode is pH sensitive

- inner glass surface has - charge
- originally, - charge is balanced
by Na+ from the glass
- ion exchange occurs because
of contact with HCl soln
H+(solution) + Na+Glass-(s)
Na+(solution) +
H+Glass-(s)
- extent of ion exchange is fixed
since [H+] is constant
- outer glass surface also has -
charge
- extent of ion exchange depends on
sample soln [H+]

- difference in extent of ion exchange
causes a potential across the membrane
- EpH electrode =
f(sample soln pH)
Ion
exchange at outer glass surface can be slow
- pH meter display may
stabilize slowly
- stirring sample soln helps